In order to achieve the highest level of sensitivity and to get rid of the coil-misbalance offset voltages a special hardware has been used and integrated into the measuring system (Model T).

Owing to improved signal-to-noise performances this measuring system reaches the sensitivity level (expressed in equivalent magnetic moment, defined as Δm= Δχ • Hac• V, Δχ is susceptibility resolution of the apparatus, Hac primary field and V volume of the sample) better than 2 nanoEMU.
We illustrate hereby the measurements on small Nb platelet performed with the latter sensitivity.

AC susceptibility measurement on Nb platelet (m=6.4 mg), taken with CryoBIND T. The measurement has been performed in the temperature sweep mode with applied (primary) magnetic field of Hac=76 mOe at frequency of 230 Hz. The region between the broken lines (doubled standard deviation) has been taken as an estimate for susceptibility resolution.

How do we know about sensitivity? Here is how: